"General failure: Only 1 out of 2 pumps responded" on ACQUITY M-Class - WKB249384
Article number: 249384
SYMPTOMS
- When attempting to trigger an acquisition on the ACQUITY M-Class, the error "General failure: Only 1 out of 2 pumps responded" appears in the console
- No other ACQUITY ICSs installed
- A power cycle does not correct the issue
ENVIRONMENT
- ACQUITY M-Class
- SCIEX OS Support Layer v1.2.1
CAUSE
Faulty or corrupted inlet method
FIX or WORKAROUND
- Delete the AcquitySettings.xml file found at C:\Program Files (x86) \ Water Instruments \ Settings
- Restart the PC
- Create a new inlet method from scratch.