Skip to main content
Waters

Poor reproducibility in Atmospheric Pressure Chemical Ionization (APCI) mode - WKB24884

Article number: 24884

SYMPTOMS

  • Poor reproducibility in APCI mode
  • Corona current fluctuates between 2 and 10 µA (set point 4 µA)

ENVIRONMENT

  • Xevo TQ-S
  • Vion IMS QToF

CAUSE

Dirty corona pin

FIX or WORKAROUND

  1. Remove the corona pin.
  2. Clean the corona pin with a mildly abrasive pad, such as Scotch Brite, under running water.
  3. Dry thoroughly.
  4. Reinstall the corona pin, ensuring that it is at the same angle as previously.

ADDITIONAL INFORMATION

 

id24884, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD

 

Not able to find a solution? Click here to request help.