CCS LockSpray setup fails on SYNAPT G2-Si - WKB24763
Article number: 24763
SYMPTOMS
- CCS LockSpray setup was passing, but it is no longer with the new LockSpray solution
ENVIRONMENT
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
CAUSE
Weak LockSpray solution
FIX or WORKAROUND
If using leucine enkephalin, use a solution that gives at least 5e5 counts per second for the lockmass.
ADDITIONAL INFORMATION
Issues may also arise if the solution is too strong (>5e6).
id24763, MALDIG2SI, SYNAPTXS, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML