How to replace the argon tank for a mass spectrometer - WKB245506
Article number: 245506
OBJECTIVE or GOAL
Replace the argon tank for a mass spectrometer.
ENVIRONMENT
- All Waters Mass Spectrometers and Time of Flight instruments
PROCEDURE
- Turn off all gases (Collision, API, TRAP, and IMS) from the Tune page.
- Close the valve on top of the argon tank.
- Close any shutoff valves between the argon tank and the instrument.
- Remove the regulator from the empty argon tank.
- Install the regulator onto the new argon tank.
- Open the valve at the top of the argon tank.
- Open any of the shutoff valves that were closed in step 3.
- Turn on collision gas from the Tune page.
ADDITIONAL INFORMATION
id245506, TQD, TQDIVD