Article number: 24512

SYMPTOMS

  • Periodic dip in trace that correlates with baffle turn

ENVIRONMENT

  • QTof Premier
  • QTof micro
  • QTof Ultima API
  • QTof Ultima Global
  • Synapt HDMS
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Vion IMS QTof
  • Xevo QTof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof

CAUSE

Dirty baffle

FIX or WORKAROUND

  1. Clean the cone and the baffle in 100% formic acid.
  2. Rinse well with water and test.

ADDITIONAL INFORMATION

If this did not work, further cleaning is required. See Chromatogram trace dips when the baffle turns on SYNAPT.