Unstable or no beam from either the ESI (sample) probe or the reference (LockSpray) probe - WKB239359
Article number: 239359
SYMPTOMS
- Unstable or no beam from either the ESI (sample) probe or the reference (LockSpray) probe
ENVIRONMENT
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- Xevo G3 QTof
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
CAUSE
Either the fluidics wash solvent is empty or the PEEK lines are not fully submerged.
FIX or WORKAROUND
- Make up fresh wash solvent.
- Submerge the PEEK lines.
- Purge fluidics.
- Confirm flow to both probes.
ADDITIONAL INFORMATION
id239359, Electrospray, MALDIG2SI, SYNG2SIHD