Skip to main content
Waters

Unstable or no beam from either the ESI (sample) probe or the reference (LockSpray) probe - WKB239359

Article number: 239359

SYMPTOMS

  • Unstable or no beam from either the ESI (sample) probe or the reference (LockSpray) probe

ENVIRONMENT

  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • Xevo G3 QTof
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS

CAUSE

Either the fluidics wash solvent is empty or the PEEK lines are not fully submerged.

FIX or WORKAROUND

  1. Make up fresh wash solvent.
  2. Submerge the PEEK lines.
  3. Purge fluidics.
  4. Confirm flow to both probes.

ADDITIONAL INFORMATION

clipboard_ef1e3d9cf865e4b3b9ec9082864e7078c.png

Not able to find a solution? Click here to request help.