Audible nitrogen leak from front of ACQUITY RDa Detector - WKB235431
Article number: 235431
SYMPTOMS
- Audible nitrogen leak from front of ACQUITY RDa Detector
- High nitrogen gas consumption
- Low desolvation gas flow error
- Source pressure test sometimes fails
ENVIRONMENT
- BioAccord System
- ACQUITY RDa Detector
CAUSE
Leaking pinch valve on RDa source exhaust
FIX or WORKAROUND
Replace the pinch valve, Part Number (700011173).