SYMPTOMS

  • Rotary pump does not start after PM
  • Oil level OK
  • Power line OK

ENVIRONMENT

  • Xevo TQD

CAUSE

The Leybold pump relay cable is faulty.

 

FIX or WORKAROUND

Replace the Leybold Pump Relay Cable (700004020).

ADDITIONAL INFORMATION

 

id23381, XEVOTQD, XEVOTQDIVD

 