RDa fails instrument setup, has low sensitity, or aborts batches with fluidics/resolution check errors - WKB229163

Article number: 229163

SYMPTOMS

  • RDa fails instrument setup
  • Low intense beam for lock mass solution (LeuEnk)
  • Instrument setup fails on detector setup and/or Autotune stage in negative mode
  • Visible ringing observed when system running detector setup
  • Fluidics Error occurs randomly
  • Resolution Check Error occurs randomly

ENVIRONMENT

  • BioAccord
  • RDa
  • waters_connect
  • UNIFI

CAUSE

Faulty capillary

FIX or WORKAROUND

Replace the capillary, ESI probe assembly 500 mm (700011871).

ADDITIONAL INFORMATION