On a SYNAPT G2, what are the maximum gas flow rates for argon, helium, and nitrogen? - WKB228617

Article number: 228617

ENVIRONMENT

  • SYNAPT G2
  • MassLynx

ANSWER

Argon has a maximum value of 10 mL/min (a typical flow rate is 2 mL/min).

Helium has a maximum value of 200 mL/min (a typical flow rate is 180 mL/min).

IMS nitrogen has a maximum value of 120 mL/min (a typical flow rate is 90 mL/min). API nitrogen gas maxes out at 1200 L/h for desolvation and 300 L/h for cone (typical flow rates vary according to method).

ADDITIONAL INFORMATION

 