Low area counts for one compound in a multi-compound assay on the TQD, Xevo TQD, Xevo TQ-S, Xevo TQ-S cronos or Xevo TQ-S Micro - WKB22757

Article number: 22757

SYMPTOMS

  • Low area counts for one compound
  • 50% of expected value
  • All other compounds show expected area counts
  • Retention time and peak integration are fine

ENVIRONMENT

  • TQD
  • Xevo TQD
  • Xevo TQ-S cronos
  • Xevo TQ-S Micro
  • MassLynx
  • Xevo TQ-S

CAUSE

Sample prep issue

FIX or WORKAROUND

  1. Prepare fresh standards.
  2. Run calibration curve.

ADDITIONAL INFORMATION