Where is the EPC reset switch located on the ToF instrument? - WKB227502
Article number: 227502
ENVIRONMENT
- Xevo G2-XS QTof
- SYNAPT XS
ANSWER
See the photo in Addtional Information.
To reset:
- Close MassLynx.
- Open the instrument front panel sliding door.
- Press the reset switch with a pin (for example, a piece of PEEK tubing).
- Wait three minutes for the EPC to restart.
- Restart MassLynx.
ADDITIONAL INFORMATION
id227502, SYNAPTXS, XEVOG2XSTF