Where is the EPC reset switch located on the ToF instrument? - WKB227502

Article number: 227502

ENVIRONMENT

  • Xevo G2-XS QTof
  • SYNAPT XS

ANSWER

See the photo in Addtional Information.

To reset:

  1. Close MassLynx.
  2. Open the instrument front panel sliding door.
  3. Press the reset switch with a pin (for example, a piece of PEEK tubing).
  4. Wait three minutes for the EPC to restart.
  5. Restart MassLynx.

ADDITIONAL INFORMATION

clipboard_e830b295abfdca5ab00320af9a76b009a.png