MassLynx displays "acquiring" but no data is acquired - WKB22330
SYMPTOMS
- The first sample injection completes as expected
- The second sample is never injected, or MS acquisition stops in the middle of the run
- In the bottom-right corner of MassLynx, it says "acquiring" but there is no scan count
- Acquisition can be stopped, but in order to start a new injection, MassLynx needs to be restarted and the MS needs to be rebooted
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-S micro
- Any MS
- MassLynx
CAUSE
Corrupted MS method.
Possible cause 1: The auto dwell for one or more functions does not work. The typical 0.005 s dwell time becomes much longer (e.g., 268435 seconds). If a function contains eight channels, it would take 596 hours to complete the first data point.
Possible cause 2: The MS method was created on a different SCN version of MassLynx.
FIX or WORKAROUND
Fix: Create a new MS method from scratch.
Workaround option 1: Use manual dwell time instead of auto dwell.
Workaround option 2: Save an older, tested version of the MS method and update it as needed.
ADDITIONAL INFORMATION
After setting a manual dwell time and then going back to auto dwell, the calculation still didn't work. The auto dwell was 268435 s again.
When the corrupted MS method was transferred to another MassLynx PC, the auto dwell worked as expected.