ENVIRONMENT
- Mass Spectrometer EPC to PC network card
- Ethernet cable UTP (unshielded twisted pair)
-
Q-Tof Premier
-
Q-Tof Ultima family
-
Q-Tof API US
-
Q-Tof 1 and 2 (TDAT)
-
Q-Tof 1 and 2 (EPCAS)
-
MALDI *
-
MALDI micro MX *
-
GCT *
-
LCT *
-
LCT Premier
-
Quattro Premier
-
AutoSpec
ANSWER
The following instruments use a crossover cable for communications between the EPC and the computer network card or internet hub:
-
Q-Tof Ultima family
-
Q-Tof Premier
-
Q-Tof API US
-
Q-Tof 1 and 2 (TDAT)
-
Q-Tof 1 and 2 (EPCAS)
-
MALDI *
-
MALDI micro MX *
-
GCT *
-
LCT *
-
LCT Premier
-
Quattro Premier
-
Quattro Premier Pt
-
AutoSpec
* Instruments with a standalone EPC outside the instrument