ADC setup fails on Vion IMS at Starting Ion Area setup for channel A - WKB207214
Article number: 207214
SYMPTOMS
- Instrument Setup hangs at ADC Setup: Starting Ion Area setup for Positive polarity, channel A
- If the setup is aborted the mass range goes from 0-1 m/z
- An electronic reboot and factory restore gets the system to the point where instrument setup can be attempted again, but it fails at the same spot
- The system has a beautiful beam in all modes, yet still fails ADC setup
ENVIRONMENT
- Vion IMS
CAUSE
No flow on the fluidics.
Instrument setup does not start the fluidics for ADC setup
FIX or WORKAROUND
- Start the LockSpray pump infusing with the baffle in reference position
- Once there is a beam start instrument setup
ADDITIONAL INFORMATION
id207214, VION