Cannot see MS TIC data in MassLynx - WKB20516
Article number: 20516
SYMPTOMS
- System injects with no errors
- PDA and Lockspray traces are visible, but the MS trace for the sample is a flat line
- Infusion from both the lockspray sprayer and the sample sprayer from MS fluidics shows signal
ENVIRONMENT
- MassLynx 4.1 SCN 884
- Xevo G2-S Tof
- ACQUITY UPLC H-Class
CAUSE
The MS method's Method Events table is not properly defined. In this case, there is an entry listing the Sample start at the Initial Condition and the Event is set to Flow State, but the Action column is blank.
FIX or WORKAROUND
Ensure that the LC is selected for Action in the Method Events table for the Sample sprayer, as shown below:
ADDITIONAL INFORMATION
id20516, MLYNX, MLYNXV41, SCN884, SUPMM