Skip to main content
Waters

ACQUITY QDa Mass Detector: Actions to take when sensitivity is poor - WKB203962

Article number: 203962

SYMPTOMS

  • Poor sensitivity of QDa detector
  • H-Class is injecting the specified injection volume

ENVIRONMENT

  • ACQUITY QDa Mass Detector
  • ACQUITY UPLC H-Class
  • Only one positive SIR channel obtained

CAUSE

  • Vacuum leak
  • Clogging of sample cone or aperture disc
  • Dirty sample cone/charge-up/aperture disc

FIX or WORKAROUND

  1. Check the turbo power.
    Check from the Console window > QDa Detector > MS display.
    37-60: Normal
    60 or more: There is a possibility of vacuum leak
    37 or less: Sample cone or aperture disc may be clogged
    clipboard_e50d6b4feaf7cb243e229378ecef64fd6.png
  2. Clean the sample cone.
    See ACQUITY QDa Detector Overview and Maintenance Guide.
  3. Add a negative SIR channel to the instrument method and set the same acquisition time with the currently existing positive SIR channel.
    Perform replicate injections using the instrument method with the above settings, and check if sensitivity drops.
  4. Replace the aperture disc.
    See ACQUITY QDa Detector Overview and Maintenance Guide.

ADDITIONAL INFORMATION

id203962, QDA, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.