Calibration fails with "Peaks too Strong" error - WKB203405
Article number: 203405
SYMPTOMS
- Cannot calibrate Xevo G2-XS
- "Peak intensity too strong" error
ENVIRONMENT
- Xevo G2-XS
CAUSE
Calibrating (in Negative mode) with sodium iodide, the lower mass peaks are very intense compared with the higher masses.
FIX or WORKAROUND
- Switch to Manual Quad Profile.
- Adjust the Quad Profile to increase the intensity of the higher mass peaks relative to the lower masses.
- Repeat the calibration.
- Switch back to Auto profile if desired.
ADDITIONAL INFORMATION
id203405, XEVOG2XSTF