Skip to main content
Waters

ACQUITY UPLC FLR Detector leaking flow cell on BioAccord System - WKB194991

Article number: 194991

SYMPTOMS

  • Leak observed from  ACQUITY UPLC FLR Detector flow cell

ENVIRONMENT

  • ACQUITY UPLC I-Class
  • ACQUITY UPLC FLR Detector
  • BioAccord System

CAUSE

High backpressure on the cell caused cell failure. Fittings checked and a kink observed in the capillary from the divert valve to source enclosure

FIX or WORKAROUND

1. Replace RDa capillary, part number: 700011871

ADDITIONAL INFORMATION

Flow applied to system at method flow rate of 0.4ml/min. Switching flow on divert valve between waste and MS showed a 1000psi increase in backpressure when flowing to the MS. FLR flow cell has a backpressure limit of 500psi

Not able to find a solution? Click here to request help.