Xevo mass spec vents shortly after sample cone is removed and replaced - WKB192807

Article number: 192807

SYMPTOMS

  • Mass spec may show a "Turbo controller fault"
  • Mass spec vents shortly after sample cone and gas cone nozzle are replaced

ENVIRONMENT

  • Xevo mass spectrometers
  • Xevo TQ-S micro
  • Xevo TQD
  • Xevo TQ-XS
  • Vion IMS QTof
  • Xevo G2 QTof/Tof
  • Xevo G2-S/XS
  • Xevo QTof MS
  • Xevo TQ-MS
  • XevoTQ-S
  • Xevo TQ-S cronos

CAUSE

The orange O-ring that sits under the gas cone nozzle on the ion block becomes worn and no longer maintains vacuum.

FIX or WORKAROUND

  1. Close the isolation valve & keep it closed

  2. Remove the cone & cone nozzle

  3. Check o-ring and replace if damaged 

  4. Re-install both cone & cone nozzle

  5. Check the vacuum

  6. Re-open the isolation valve

If these steps don't resolve the problem

Replace the orange O-ring (700004728 for a package of 5)