Regular drop in signal (0.4 min) with SYNAPT G2-Si - WKB18266
Article number: 18266
SYMPTOMS
- Regular signal drop spikes observed on chromatogram every 0.4 minutes
ENVIRONMENT
- SYNAPT G2-Si
- SYNAPT G2-S
- MassLynx
CAUSE
Trap and Transfer vacuum gauges on the vacuum tab of the tune page are seen to drop and recover with signal drop.
The issues stems from a faulty argon supply regulator, which is not able to deliver stable outlet pressure of 7 psi.
FIX or WORKAROUND
Replace the regulator.
ADDITIONAL INFORMATION