Scroll pump does not turn off after venting MS - WKB1657
Article number: 1657
SYMPTOMS
- Software says system is vented, but the pumps are still running
ENVIRONMENT
- Synapt G2-Si with Edwards XDS35i and XDS100B scroll pumps
- Xevo TQ-S
CAUSE
The covered on/off switch on the base is set to "I" (always on).
FIX or WORKAROUND
- Switch to the other position (software control).
ADDITIONAL INFORMATION
Switch location: