The capillary is being pushed out of the probe - WKB1299

Article number: 1299

SYMPTOMS

  • High system pressure
  • No MS beam
  • The capillary is being pushed out of the probe

ENVIRONMENT

  • Quattro Micro
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Xevo G2-S
  • Xevo G2-XS

CAUSE

  • Failure to flush buffers out of the system before pumping organic solvent
  • Salt precipitates out of solution when organic solvents hit them, causing restriction in the capillary
  • Restriction from salt precipitates causes capillary to be pushed out of position

FIX or WORKAROUND

  1. Rebuild the probe
  2. Before you connect to a newly rebuilt probe, flush buffers out of the LC system and route to waste