The capillary is being pushed out of the probe - WKB1299
Article number: 1299
SYMPTOMS
- High system pressure
- No MS beam
- The capillary is being pushed out of the probe
ENVIRONMENT
- Quattro Micro
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo G2-S
- Xevo G2-XS
CAUSE
- Failure to flush buffers out of the system before pumping organic solvent
- Salt precipitates out of solution when organic solvents hit them, causing restriction in the capillary
- Restriction from salt precipitates causes capillary to be pushed out of position
FIX or WORKAROUND
- Rebuild the probe
- Before you connect to a newly rebuilt probe, flush buffers out of the LC system and route to waste
ADDITIONAL INFORMATION
id1299, QUMICRO