System status in MassLynx shows "Inlet Not Ready" - WKB124755
Article number: 124755
SYMPTOMS
- System Status on MassLynx main page shows "Inlet Not Ready"
- Status in the Inlet Method shows Ready as green and OK as red
- No error message shows the ACQUITY MS Console for any of the individual components of the system
- LC > Reset communications from the Inlet method editor does not resolve
- Double-click on red OK LED shows communication error with ACQUITY UPLC Column Manager
ENVIRONMENT
- ACQUITY UPLC H-Class
- MassLynx 4.2
CAUSE
Communication issue between ACQUITY UPLC Column Manager and MassLynx
FIX or WORKAROUND
- Open the ACQUITY MS Console.
- Select the Column Manager in the System list on the left-hand side.
- From the Control Menu, select the RESET CM option.
ADDITIONAL INFORMATION
Note: In the ACQUITY MS Console, in the Control menu for each of the individual components of the system, there is a RESET option that can be used to do a soft reset of each component.