What is the TOF Vacuum Operate Trip level for SYNAPT G2-Si and Xevo G2-XS? - WKB12425
Article number: 12425
ENVIRONMENT
- Synapt G2-Si
- Synapt G2-S
- Synapt G2
- Xevo G2-XS
- Xevo G2-S
ANSWER
The instrument cannot be switched into Operate unless the TOF pressure is below 1.2E-6 mbar and all of the turbo pump speeds are greater than 80%.
ADDITIONAL INFORMATION
For Xevo G2 QTof and Xevo QTof, please see: What is the TOF Vacuum Operate Trip level for Xevo G2 Qtof and Xevo Qtof?