How to replace the outer source heaters on the AutoSpec - WKB118407
Article number: 118407
OBJECTIVE or GOAL
Replace the outer source heaters on the AutoSpec.
ENVIRONMENT
- AutoSpec Premier
- AutoSpec Ultima
PROCEDURE
See the attachment.
ADDITIONAL INFORMATION
- Recommendation: Wear gloves that have been washed with pure water.
- Position the upper heater loop counterclockwise, so upon tightening the heater it is not dragged out of position.
- Before returning the outer source assembly into the instrument, measure the heater resistances (1 or 2 Ohms expected) and the resistance between the outer source block and the Focus 1 lens (open circuit expected).