Quattro Premier XE mass spectrometer does not respond to MassLynx commands - WKB11046
Article number: 11046
SYMPTOMS
- Cannot put mass spectrometer into standby
- Cannot turn off the gases
- The mass spec status in MassLynx is "Instrument Present"
- When a batch is submitted, mass spec status does not change to "Waiting for inlet start 1"
ENVIRONMENT
- MassLynx 4.1
- Quattro Premier XE
- Xevo TQD
- Xevo TQ-S micro
CAUSE
No communication between MassLynx and the Quattro Premier
FIX or WORKAROUND
- Reboot the EPC:
- Close MassLynx.
- Go to the Windows START button and select Run from the menu. (If Run is not available, press the Windows log and R keys on the keyboard simultaneously.)
- Specify telnet epc in the Run dialog.
- In the Telnet window, at the blinking cursor, type reboot and hit Enter.
- If the Telnet window does not open, push the reset button in the larger (bottom) hole on the lower-right corner of the front of the instrument.
- Wait about two to thyree minutes for epc to reboot.
- Open MassLynx.
ADDITIONAL INFORMATION
id11046, comms, MLYNX, MLYNXV41, QUATPREMXE, QUPREMIER, SUPMM, XEVOTQD, XEVOTQDIVD