How can the microscope slide be secured into the slide plate on a MALDI SYNAPT G2? - WKB10440
Article number: 10440
ENVIRONMENT
- MALDI
- SYNAPT G2
- HDI (high-definition imaging)
- Microscope glass slide
- Microscope slide plate (700003896)
ANSWER
The glass slide is held in place with a spring clip, located at the top of the MALDI plate. This spring may require adjusting to apply enough pressure to the microscope slide.
Double-sided tape can be used between the slide and the plate (it does not need to be conductive). Do not apply the tape so that it covers the entire slide; use two small rectangles (~10 mm x 5 mm) on the left- and right-hand sides of the glass slide. If possible, avoid placing tape behind the tissue section.
ADDITIONAL INFORMATION
id10440, MALDIG2SI, MALDIMICRO, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL