MS diverter valve does not switch to waste - WKB1041
Article number: 1041
SYMPTOMS
- The diverter valve is scheduled to turn to waste at X minutes, but it does not
ENVIRONMENT
- MassLynx 4.1
- Xevo MS (quads and QTofs)
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
CAUSE
The diverter valve is being asked to switch at the end of the run time. That is, the run time = the time when the diverter valve is being asked to move. The run is over, so the software performs no more actions.
FIX or WORKAROUND
Change the run time to a point later than the diverter valve switch time (e.g., run time = X.1 minutes and the divert time = X minutes).
ADDITIONAL INFORMATION
id1041, MALDIG2SI, MLYNX, MLYNXV41, SUPMM, SYNG2SIHD