Source pressure test failing due to source leak - WKB100925

Article number: 100925

SYMPTOMS

  • Source pressure test fails.
  • Console gives error "Source pressure test leak warning".
  • Customer recently cleaned the sample cone.

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQ-MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

O-ring between the sample cone and the gas cone nozzle is damaged or seated improperly.

FIX or WORKAROUND

Replace O-ring behind sample cone, 700004321.

ADDITIONAL INFORMATION

 

 