Sample temperature range error on ACQUITY sample manager - WKB99950
Article number: 99950
SYMPTOMS
- Sample temperature range error (50) when you select 5 ºC
ENVIRONMENT
- ACQUITY Sample Manager
- SM-FTN
CAUSE
Windows regional settings are not configured for English (United States).
The symbols for English (United States) use decimal formatting, not comma formatting, which is important when
displaying numeric values (12.56, not 12,56).
FIX or WORKAROUND
Configure Windows regional settings for English (United States).
ADDITIONAL INFORMATION