Sample temperature range error on ACQUITY sample manager - WKB99950

Article number: 99950

SYMPTOMS

  • Sample temperature range error (50) when you select 5 ºC

ENVIRONMENT

  • ACQUITY Sample Manager
  • SM-FTN

CAUSE

Windows regional settings are not configured for English (United States).

The symbols for English (United States) use decimal formatting, not comma formatting, which is important when
displaying numeric values (12.56, not 12,56).

ADDITIONAL INFORMATION

 