Alliance 2695: The mobile phase/solvent in the bottle runs out or decreases quickly even though it’s not flowing - WKB99716
Article number: 99716
SYMPTOMS
- The mobile phase/solvent in the bottle runs out or decreases quickly even though it’s not flowing
ENVIRONMENT
- 2695 Separations Module
CAUSE
GPV (gradient proportioning valve) inlet/outlet tubing is loose and leaking
FIX or WORKAROUND
Tighten the screws (red ● in the figure below) of each tube on the gradient proportioning valve.
ADDITIONAL INFORMATION
To access the GPV, open the right hand panel on the solvent manager and gently pull out the solvent tray to inspect the connections.