Alliance 2695: The mobile phase/solvent in the bottle runs out or decreases quickly even though it’s not flowing - WKB99716

Article number: 99716

SYMPTOMS

  • The mobile phase/solvent in the bottle runs out or decreases quickly even though it’s not flowing

ENVIRONMENT

  • 2695 Separations Module

CAUSE

GPV (gradient proportioning valve) inlet/outlet tubing is loose and leaking

FIX or WORKAROUND

Tighten the screws (red ● in the figure below) of each tube on the gradient proportioning valve.

clipboard_e9c679da875356406b3ef102a4c026c9d.png

ADDITIONAL INFORMATION

To access the GPV, open the right hand panel on the solvent manager and gently pull out the solvent tray to inspect the connections.

 

clipboard_e23c65dfefc2f14d913b5e396249fcc46.png