Low pressure with 2695 SM - WKB99641

Article number: 99641

SYMPTOMS

  • Low pressure readback (<50 psi) when solvent flowing
  • Passes static leak test

ENVIRONMENT

  • 2695 Separations Module

CAUSE

Leak in sample loop

FIX or WORKAROUND

Replace the sample loop. 

ADDITIONAL INFORMATION

WAT270981 - Sample Loop, 100-µL, Standard (Phase I)

430001194 - Sample Loop, 100-µL, Standard (Phase II)

 

 