ACQUITY Arc system reports instrument failure when configured with a CM-30S column heater - WKB99300
Article number: 99300
SYMPTOMS
- ACQUITY Arc system was upgraded from the originally configured CH/C column heater to a CM-30S column heater
- When column 7 and column 8 are set in the instrument method, the instrument reports "instrument failure" during injection and instrument communication fails
- The LAC/E must be restarted to rebuild the connection
ENVIRONMENT
- ACQUITY Arc
- CM-30S
CAUSE
Incompatible column heater driver
FIX or WORKAROUND
- Uninstall all the existing driver packs on LAC/E.
- Install ACQUITY CM 1.61 in the UPC2 ACQUITY 2014 Driver Pack.
- Install other drivers in the latest driver pack (2019R1), and choose not to upgrade the column heater driver during the installation process.