2695 Separations Module: What are the considerations when cleaning with 6N nitric acid for passivation treatment? - WKB99286

Article number: 99286

ENVIRONMENT

  • 2695 Separations Module

ANSWER

- It seems that the solvent filter (the filter attached to the end of the tube going into the solvent bottle) is often removed during cleaning and subsequent measurements.
- There is no fixed grade of nitric acid, but it seems that the First grade or Special grade is ideal.
- For cleaning with water after cleaning with nitric acid, confirm that it has become neutral with Litmus paper.

ADDITIONAL INFORMATION

 

id99286, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM

Not able to find a solution? Click here to request help.