Liquid found on top of vials after injection on 2695 SM - WKB99094

Article number: 99094

SYMPTOMS

  • Liquid appears on top of vials in the sample chamber

ENVIRONMENT

  • Alliance HPLC system
  • 2695 Separations Module

CAUSE

Seals are not adjusted (needle step motor could lose steps over time).

Seal pack requires replacement 

FIX or WORKAROUND

  1. Perform seal adjustment.
  2. Repeat injection and verify correct system functionality, if not functioning still replace the seal pack part number WAT271019 for standard and 700011783 for performance plus rebuild kit. 

ADDITIONAL INFORMATION