Liquid found on top of vials after injection on 2695 SM - WKB99094
Article number: 99094
SYMPTOMS
- Liquid appears on top of vials in the sample chamber
ENVIRONMENT
- Alliance HPLC system
- 2695 Separations Module
CAUSE
Seals are not adjusted (needle step motor could lose steps over time).
Seal pack requires replacement
FIX or WORKAROUND
- Perform seal adjustment.
- Repeat injection and verify correct system functionality, if not functioning still rebuild the seal pack WAT271019 and if that does not fix the issue, replace the seal pack part number 700002791 for standard and 700011783 for performance plus rebuild kit.
ADDITIONAL INFORMATION
