"Pump H/W Homing Error (3)" with ACQUITY UPLC H-Class QSM - WKB98576
Article number: 98576
SYMPTOMS
- Pump H/W Homing Error (3) occurs during initialization
- Removing pump head brings successful initialization of motor
ENVIRONMENT
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class
CAUSE
Restriction of accumulator pump head movement due to resistance from back pressure or physical restriction
FIX or WORKAROUND
- Remove the pump head and confirm free travel.
- Clean the plunger assembly, head, and pump springs to remove buildup of precipitates.
- Reinitialize the instrument to verify that the issue is resolved.