ACQUITY QSM Error: "[Primary or Accumulator] Unable to compress" - WKB98287
Article number: 98287
SYMPTOMS
Quaternary Solvent Manager (QSM) Dynamic leak test failing with "Primary unable to compress" error.
Quaternary Solvent Manager (QSM) Dynamic leak test failing with "Accumulator unable to compress" error.
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
CAUSE
Seals unable to hold pressure. Rebuild pump head.
FIX or WORKAROUND
- Perform pump head maintenance
ADDITIONAL INFORMATION
- Can replace parts individually
- Or use a pumphead rebuild kit, which include the appropriate check valve: