2695 Separations Module Sample Heater/Chiller will not cool to set point - WKB97064

Article number: 97064

SYMPTOMS

2695 Separations Module Sample Heater/Chiller will not cool to set point.

ENVIRONMENT

  • 2695 Separations Module
  • Alliance HPLC System

CAUSE

Condensation has become frozen in Sample Chiller engine.

FIX or WORKAROUND

Run the Defrost function.

ADDITIONAL INFORMATION

2695 Separations Module Operator's Guide see section 8.4.12 for defrost instruction

id97064, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM