2695 Separations Module Sample Heater/Chiller will not cool to set point - WKB97064
Article number: 97064
SYMPTOMS
2695 Separations Module Sample Heater/Chiller will not cool to set point.
ENVIRONMENT
- 2695 Separations Module
- Alliance HPLC System
CAUSE
Condensation has become frozen in Sample Chiller engine.
FIX or WORKAROUND
Run the Defrost function.
ADDITIONAL INFORMATION
2695 Separations Module Operator's Guide see section 8.4.12 for defrost instruction
id97064, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM