Liquid on top of vials and carousels in sample compartment of 2695 Separations Module - WKB96838
Article number: 96838
SYMPTOMS
- Needle wash solvent leaking into sample compartment
- Leak observed after injections and needle wash prime
- Liquid on top of vials and carousels
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
CAUSE
The yellow needle wash waste line is kinked, blocked, or submerged in liquid waste.
FIX or WORKAROUND
- Pull the yellow needle wash line out of liquid waste.
- Smooth out any kinks.
- If necessary, replace or shorten the needle wash waste line (WAT077040).
- Verify that needle wash prime works.
ADDITIONAL INFORMATION
id96838, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, wet