Column Manager - Active: No temperature response from the active preheater (APH) in console - WKB9662
Article number: 9662
SYMPTOMS
- No temperature response from the Active preheater (APH) in console
- CM-A/CM-Aux configuration:
- Four active preheaters in the four column banks; none respond to temperature settings per the console
- Set column 4 to 50 °C; the column temperature goes up, but according to the plots, the preheater remains at room temperature
- Console shows them listed as "present"
- APHs are installed in the correct column positions
- APH positions are all configured as "Active" in console
- Instrument method control has Active Preheater selected as "Use Console Configuration"
- After removing the APH and setting a temperature, the CM-A does not trigger the expected "preheater is configured but not detected" message
ENVIRONMENT
- ACQUITY UPLC Column Manager - Active.
- ACQUITY UPLC Auxiliary Column Manager - Active.
CAUSE
The sensor is not recognized because the latch is broken and not securely sealing the cover onto the trough.
FIX or WORKAROUND
- Replace the "Latch Set" or "Column Module Front Trough Cover".
- 700005980 for the LATCH SET, CM-A/CM-AUX TROUGH COVER -L&R
- 700005439 for the Column Module Front Trough Cover
ADDITIONAL INFORMATION
- Closing the compartment trough cover engages a magnet switch that turns on the 24V to the Active preheater.
- The active preheater will operate at the set-point temperature of the selected column position when the trough compartment cover is closed.
id9662, A-30MCM, UPC2CMAUX, UPCMA, UPCMAIVD, UPCMAUX, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPCMA, UPPCMAUX, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO