How to switch between CH-A and 30cm column heater with the same system configuration (with column switch box) - WKB96308
Article number: 96308
OBJECTIVE or GOAL
Switch between CH-A and 30cm column heater.
ENVIRONMENT
- ACQUITY UPLC H-Class Sample Manager FTN
- ACQUITY UPLC I-Class Sample Manager FTN
- ACQUITY UPLC Column Heater (30 cm)
- Column switch box (205001802)
PROCEDURE
- Open the Console screen.
- Select Sample Manager FTN from the tree on the left.
- Select Column Manager Configuration from the Configure menu.
- Select Column Module in Fluidics Configuration.
- Select [CH-A] or [CH30-A] from the pull-down menu.
ADDITIONAL INFORMATION
id96308, A-30HCM, A-30SM, UPARBCH30, UPCH30, UPCH30A, UPCHA, UPCHAIVD, UPISMFTN, UPPCHA, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO