Peak tailing with ACQUITY UPLC and H-Class systems - WKB95810

Article number: 95810

SYMPTOMS

  • Peak tailing observed in chromatograms on both ACQUITY UPLC and H-Class systems
  • New mobile phases did not rectify issue
  • Solvent manager and sample manager diagnostics pass
  • No column pre-filters used
  • No tubing modifications or maintenance performed on systems

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC H-Class

CAUSE

Column failure

FIX or WORKAROUND

Replace the column.

ADDITIONAL INFORMATION

 

id95810, eluent, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPBINARY, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.