Two-column regeneration valve does not work - WKB95711
Article number: 95711
SYMPTOMS
- Column position 1 or 2 is set to change in the "event", but there is no action
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2795 Separations Module
- Alliance Series Column Heater
- Alliance Series Column Heater/Cooler
CAUSE
The function of "Inject Immediate Samples" (Inject Immediate Standards) or "Condition Column" in the sequence cannot trigger the change of valve position in "event".
FIX or WORKAROUND
Use the "Inject Samples" or "Inject Standards" function to trigger the change of valve position.
ADDITIONAL INFORMATION
