Sample remains after injection on 2767 - WKB95500
Article number: 95500
SYMPTOMS
- The volume of aspirated sample becomes small and sample remains with 2767
- Air entrainment in the tube
ENVIRONMENT
- 2767 sample manager
CAUSE
Loose screw on tube or defective mini wash pump or tube
FIX or WORKAROUND
1. Retighten the screw on the tube.
2. If step 1 does not work, replace the mini wash pump (700001311 - PUMP, UNIVERSAL MINIWASH, ASSY).
3. If the issue persists, replace the thin tubing of T6, T7, and T8 to fix the issue.