Sample remains after injection on 2767 - WKB95500

Article number: 95500

SYMPTOMS

  • The volume of aspirated sample becomes small and sample remains with 2767
  • Air entrainment in the tube

ENVIRONMENT

  • 2767 sample manager

CAUSE

Loose screw on tube or defective mini wash pump or tube

FIX or WORKAROUND

1. Retighten the screw on the tube.

2. If step 1 does not work, replace the mini wash pump (700001311 - PUMP, UNIVERSAL MINIWASH, ASSY).

3. If the issue persists, replace the thin tubing of T6, T7, and T8 to fix the issue.

ADDITIONAL INFORMATION

id95500, 2767, 2767SFC