On ACQUITY systems, pressure is high with column or column inline filter, normal without column - WKB95443

Article number: 95443

SYMPTOMS

  • High or erratic system pressure with column in place, with or without inline filter
  • Pressure is normal with column removed

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC Column Heater
  • ACQUITY APC
  • ACQUITY ARC

CAUSE

Blocked frit

FIX or WORKAROUND

  1. Replace frit at the column inline filter or column inlet.

ADDITIONAL INFORMATION

Installation instructions: https://help.waters.com/help/en/support/library-details.html?documentid=716001783