How do I install the PATROL Corflags.exe for Windows 10? - WKB94953

Article number: 94953

OBJECTIVE or GOAL

To install the PATROL Corflags.exe for Windows 10.

ENVIRONMENT

PATROL UPLC Process Analysis System

PROCEDURE

  1. Download and install the latest “.NET Framework Developer Pack" from the following link: https://dotnet.microsoft.com/download/visual-studio-sdks. It is located under the unsupported section as of August 2024.
  2. Once installed, search the PC for the corflags.exe file. 
  3. Navigate to the corflags.exe file and open a command prompt window. In this example, the file was in the C:\Program Files (86)\Microsoft SDKs\Windows\v10.0A\bin\netfx 4.7.2 tools folder.
  4. Drag and drop the corflags.exe file into the command prompt window. The prompt should look like this (file location and executable):
    “C:\Program Files (86)\Microsoft SDKs\Windows\v10.0A\bin\netfx 4.7.2 tools\Corflags.exe”
  5. At the end of the text already in the window, manually type: /32bitreq+ c:\empower\instruments\bin\patrol.scr.  
  6. The command line should look like this:
    “C:\Program Files (86)\Microsoft SDKs\Windows\v10.0A\bin\netfx 4.7.2 tools\Corflags.exe” /32bitreq+ c:\empower\instruments\bin\patrol.scr
    (must have a space after Corflags.exe” and after 32bitreq+).
  7. Press Enter to initiate to command and set the 32-bit flag.

clipboard_e6b8d16e12e5ad49ad4d5f1c86bcba70c.png

If the PATROL software is reloaded, repeat this procedure.

ADDITIONAL INFORMATION

Execution of these instructions will result in a change to the PATROL.scr file being noted in Verify Files. e.g.

Waters Installation Qualification - Summary

The following changes were detected:

  7: PATROL.scr             CHANGED ! Size:   20992, CRC: 35187

Not able to find a solution? Click here to request help.