With ACQUITY UPLC, abnormal peak shape when analyzing naphthalene - WKB94377
Article number: 94377
SYMPTOMS
- When naphthalene is analyzed, it is okay if the eluent is methanol or acetonitrile, but the peak shape becomes abnormal when it is in THF
- All leak tests pass
- Mobile phase feed is normal with no pulsation
- The peak shape is abnormal even when the sample injection volume is 2 μL and 10 μL
- The sample injection method is a partial loop
- The weak wash solvent is the same as the initial gradient conditions
- The same THF is used as the sample dilution solvent
- The peak shape is normal when measured under the same measurement conditions, the same sample, and the same column with another ACQUITY UPLC H-Class
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC PDA eλ Detector
- SQ Detector
CAUSE
Without using the column stabilizer, port 6 of the injector valve to the column was connected with a PEEK tube.
FIX or WORKAROUND
Analyze using a column stabilizer (205000291 for 50-100 mm columns).
ADDITIONAL INFORMATION
id94377, A-30PDA, A-SQ, SQD, UPAPDA, UPARPPDA, UPBSMIVD, UPC2PDA, UPIBSMIVD, UPLCCOLHTR, UPNBINARY, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA, UPSMIVD, UPSMPMGR