Poor reproducibility on H-Class SM-FTN - WKB93273
Article number: 93273
SYMPTOMS
- High %RSD on a customer method.
- All system diagnostic tests pass, including Needle Seal Readiness test.
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
CAUSE
Leak around needle and needle seat causing high %RSD.
FIX or WORKAROUND
- Replace needle, needle seat, and needle guide.
- Part numbers (Needle, 15uL SS w/Guide and Seat)
- H-Class-700005215
- H-Class Bio-700005421
- H-Class Plus-700011628
- H-Class Bio Plus-700011658
- Characterize the needle z axis.
- Characterize the needle seal.
- Perform the Needle Seal Readiness test.
ADDITIONAL INFORMATION
id93273, A-30SM, UPBSMIVD, UPIBSMIVD, UPISMFTN, UPLCCOLHTR, UPNBINARY, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR