Error: Column Active Preheater configured but not detected in the ACQUITY Console - WKB92512
Article number: 92512
SYMPTOMS
- Error: "Column Active Preheater configured but not detected in the ACQUITY Console" appears after the column is switched from CH-A to the 30-cm CH.
- Resetting the FTN does not resolve this.
- Restarting the FTN does not resolve this.
- Reseating the Active Pre Heater (APH) does not resolve this.
ENVIRONMENT
- ACQUITY UPLC H-Class PLUS
- CH-A
- 30-cm Column Heater
- APH
CAUSE
Configuration in the Console does not match physical configuration.
FIX or WORKAROUND
- In the ACQUITY Console, click FTN.
- Click Configure.
- Click Column Module.
- Under the Column Module Configuration, use the drop-down menu and switch to CH-30A (see screenshot below).
- Click OK.
- Reset the FTN.
- Replace the APH . Part number 205000730
ADDITIONAL INFORMATION
In UNIFI 1.9.4 a wrong column manager configuration can be pulled from the startup/shutdown method and also from the the system startup procedure. Make sure you use the correct configuration
id92512, A-30HCM, UPCHA, UPCHAIVD, UPPCHA, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO